DocumentCode :
2879287
Title :
The PI-method: non-redundant data capture and efficient reconstruction for helical cone-beam CT
Author :
Turbell, Henrik ; Danielsson, Per-Erik
Author_Institution :
Dept. of Electr. Eng., Linkoping Univ., Sweden
Volume :
3
fYear :
1998
fDate :
1998
Firstpage :
1424
Abstract :
Helical cone-beam scanning is a technique for rapidly acquiring X-ray projections of a long object. The authors present a detector shape which ensures complete and non-redundant data capture. Furthermore, a recently invented, computationally efficient reconstruction algorithm, produces images of high quality
Keywords :
computerised tomography; image reconstruction; medical image processing; PI-method; X-ray projections; computationally efficient reconstruction algorithm; efficient reconstruction; helical cone-beam CT; high quality images; long object; medical diagnostic imaging; nonredundant data capture; Computed tomography; Detectors; Fans; Image processing; Image reconstruction; Laboratories; Lighting; Object detection; Reconstruction algorithms; X-ray imaging;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium, 1998. Conference Record. 1998 IEEE
Conference_Location :
Toronto, Ont.
ISSN :
1082-3654
Print_ISBN :
0-7803-5021-9
Type :
conf
DOI :
10.1109/NSSMIC.1998.771653
Filename :
771653
Link To Document :
بازگشت