Speckle reduction in two easy steps
Author
Leeman, S. ; Gatenby, J.C. ; Forsberg, F. ; Hoddinott, J.C.
Author_Institution
King´´s Coll. Sch. of Med. & Dentistry, London, UK
fYear
1989
fDate
3-6 Oct 1989
Firstpage
927
Abstract
A somewhat unorthodox technique for speckle reduction in medical ultrasound pulse-echo images is discussed. Two basic difficulties must be overcome. Recognition of the artifact is based upon an analysis of the instantaneous (temporal) frequency of the individual RF A-line. Correction of the speckle artifact at the local regions indicated by the recognition procedure is performed via techniques derived from the frequency-diversity approach. The hybrid temporal-Fourier processing algorithm has the merit of ultimate applicability on a real-time basis
Keywords
acoustic imaging; biomedical ultrasonics; patient diagnosis; picture processing; speckle; RF A-line; frequency-diversity approach; hybrid temporal-Fourier processing algorithm; instantaneous temporal frequency; medical ultrasound pulse-echo images; real-time processing; speckle reduction technique; Biomedical imaging; Frequency diversity; Image analysis; Image recognition; Pulse measurements; Pulse shaping methods; Scattering; Signal processing; Speckle; Ultrasonic imaging;