DocumentCode
2901574
Title
Software adjustment of pressure transducers
Author
Opris, Ioana ; Blaga, Petre
Author_Institution
Univ. Politeh. of Bucharest, Bucharest, Romania
fYear
2015
fDate
7-9 May 2015
Firstpage
9
Lastpage
14
Abstract
The calibration procedure of a pressure transducer changed over time along with the development of transducers from analog to smart. The paper presents the linear regression as a method for verification and software adjustment of the digital output of hybrid and smart pressure transducers. The method consists in using the comparison regression line for determining the adjusting regression line and the adjustment factors of the digital output signal. The method is simple, efficient, considering all together the known and unknown influences which alter the measurements. A practical example for the verification and adjustment of a pressure transducer is presented in the paper.
Keywords
calibration; intelligent sensors; pressure transducers; calibration; digital output; linear regression; pressure transducers; software adjustment; verification; Calibration; Linear regression; Microprocessors; Software; Standards; Transducers; Voltage measurement; linear regression; pressure transducer; software adjustment;