DocumentCode :
2904123
Title :
Six-Port Reflectometer: Influence of Q-Points Position in r-Plane on Sidearm Power Detector Error Propagation
Author :
Brunetti, L. ; Fornero, C. ; Rietto, G.
Author_Institution :
METF Division, IEN Galileo Ferraris
fYear :
1988
fDate :
7-10 June 1988
Firstpage :
274
Lastpage :
275
Abstract :
An attractive feature of the network analyzers based on six-port measurement technique is the high number of equivalent multiport junctions that at least theoretically may be implemented. The paper describes an effort to look for an optimum circuit configuration that minimizes the propagation of the errors due to power detectors.
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements, 1988. CPEM 88 Digest. 1988 Conference on
Conference_Location :
Ibaraki, Japan
Type :
conf
DOI :
10.1109/CPEM.1988.671293
Filename :
671293