Title :
3D ultrasound imaging system using Fresnel ring array
Author :
Tamano, Satoshi ; Yamazaki, Masashi ; Sano, Shuzou ; Hara, Kenji ; Sakano, Junichi ; Miwa, Yuuichi
Author_Institution :
R&D Center, Hitachi Med. Corp., Tokyo, Japan
Abstract :
A real-time 3D ultrasound system using Fresnel ring array is proposed. We have developed a prototype of a convex-convex shaped 2D-array probe with curvature of 40 mm for both in the lateral-axis direction and the elevational (vertical)-axis direction, and the center frequency is 3.5MHz. As a result, a transmitted beam profile measured in water is in good accordance with the theoretical simulation using FEM. By using the convex-convex shaped 2D array probe including a newly developed multiplex-IC, the dynamic focusing is performed with Fresnel ring, so that a good isotropic beam for scan can be obtained. A preliminary real-time 3D image has been demonstrated.
Keywords :
biomedical electronics; biomedical ultrasonics; finite element analysis; ultrasonic arrays; ultrasonic imaging; 3.5 MHz; 3D ultrasound imaging system; FEM; Fresnel ring array; convex-convex shaped 2D array probe; dynamic focusing; elevational axis; isotropic beam; lateral axis; multiplex-IC; real-time 3D image; transmitted beam profile; Acoustic beams; Biomedical imaging; Circuits; Focusing; Frequency; Image quality; Probes; Prototypes; Real time systems; Ultrasonic imaging;
Conference_Titel :
Ultrasonics, 2003 IEEE Symposium on
Print_ISBN :
0-7803-7922-5
DOI :
10.1109/ULTSYM.2003.1293143