Title :
Influence of malfunctioning block detectors on the calculation of single detector efficiencies in PET
Author :
Ferreira, Nuno C. ; Trébossen, R. ; Grégoire, M.C. ; Bendriem, B.
Author_Institution :
Service Hosp. Frederic Joliot, Orsay, France
Abstract :
The authors studied the artifacts generated by the application of methods of crystal efficiency calculation based on the fan-sum, to situations of high variability of efficiency, namely the cases of a ring with defective blocks and acquisitions using narrow energy windows. The authors propose two efficiency calculation methods to reduce these artifacts: one method discriminates the crystals with an efficiency very different from the average, and then uses a modification of the fan-sum to estimate the efficiency; the other method calculates iteratively the fan-sum of each crystal, weighted by the efficiencies of the opposed detectors. The effect of using an off-center uniform cylinder is also discussed, since it influences the performance of the algorithms. The proposed algorithms are compared with the more conventional fan-sum method and with an algorithm used in commercial tomograph, using measurements and simulations. The results show a significant reduction of the efficiency estimation errors, in severe situations where detectors in a ring have very different detection efficiency
Keywords :
biomedical equipment; gamma-ray detection; positron emission tomography; PET; algorithms; artifacts; detector ring; efficiency calculation methods; fan-sum; iterative calculation; malfunctioning block detectors; medical diagnostic imaging; nuclear medicine; single detector efficiencies calculation; Detectors; Estimation error; Geometry; Iterative algorithms; Light scattering; Particle scattering; Photomultipliers; Photonic crystals; Positron emission tomography; Sampling methods;
Conference_Titel :
Nuclear Science Symposium, 1998. Conference Record. 1998 IEEE
Conference_Location :
Toronto, Ont.
Print_ISBN :
0-7803-5021-9
DOI :
10.1109/NSSMIC.1998.773819