Abstract :
Experimental measurements of a point source in a uniformly-attenuating cylindrical phantom are being used to model the energy-dependent point-source response function (PSRF) of a SPECT imaging system. These data provide a model PSRF for validating Monte Carlo simulations and testing reconstruction algorithms. Additionally, the data are being used to develop a reconstruction method that utilizes full energy spectral information. The experimental image data have been fitted in two stages to produce a concise description of the system PSRF. The PSRF, as a function of energy and spatial location, is specified by 14 parameters, modeling the image of a point source that would be measured at that location. The PSRF of the SPECT system is being parameterized further by modeling each of the 14 parameters by analytic functions of energy and spatial location. In this way, the energy-dependent PSRF of the SPECT imaging system can be calculated at any spatial location and energy. This paper reports the results of the first stage parameterization, the results of image reconstruction based on the first stage parameterization, and the status of the second stage parameterization of the measured system PSRF
Keywords :
Monte Carlo methods; curve fitting; image reconstruction; medical image processing; modelling; single photon emission computed tomography; Monte Carlo simulations validation; Tc; Tc-99m SPECT imaging; experimental image data; full energy spectral information; measured energy-dependent point-source response function; medical diagnostic imaging; multidimensional curve fitting; nuclear medicine; point source image modeling; reconstruction algorithms testing; spatial location; uniformly-attenuating cylindrical phantom; Biomedical imaging; Curve fitting; Energy measurement; Fasteners; Image analysis; Imaging phantoms; Multidimensional systems; Position measurement; Reconstruction algorithms; Testing;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium, 1998. Conference Record. 1998 IEEE
Conference_Location :
Toronto, Ont.
ISSN :
1082-3654
Print_ISBN :
0-7803-5021-9
Type :
conf
DOI :
10.1109/NSSMIC.1998.773854
Filename :
773854
Link To Document :
بازگشت