• DocumentCode
    2911197
  • Title

    Attenuation and scatter correction of list-mode data driven iterative and analytic image reconstruction algorithms for rotating 3D PET systems

  • Author

    Reader, A.J. ; Erlandsson, K. ; Ott, R.J. ; Flower, M.A.

  • Author_Institution
    R. Marsden NHS Trust, Inst. of Cancer Res., Sutton, UK
  • Volume
    3
  • fYear
    1998
  • fDate
    1998
  • Firstpage
    1976
  • Abstract
    Image reconstruction directly from list-mode data requires different data correction techniques to standard projection-data based reconstruction, particularly in the case of iterative reconstruction. Attenuation and scatter correction techniques have been developed for two list-mode data driven reconstruction algorithms (FAIR-B (iterative) and Atrax (analytic/iterative)) recently proposed by the authors, and the results compared with two popular projection-data based algorithms (FORE+OSEM (iterative) and FORE+FBP (analytic)). List-mode data driven algorithms require event-by-event correction schemes, or alternatively image space procedures, as no direct operations to the completely sampled projection data set can be practically carried out. The methods developed in this work allow correction of list-mode data driven EM-ML type algorithms, such as FAIR, as well as analytic list-mode algorithms. The correction schemes have been applied to simulated data from various activity and attenuating medium distributions for a rotating 3D PET system. Both list-mode algorithms show improvements over the projection-data based methods in some situations
  • Keywords
    Bayes methods; convolution; image reconstruction; iterative methods; maximum likelihood estimation; medical image processing; positron emission tomography; Atrax algorithm; Bayesian regularisation; EM-ML type algorithms; FAIR-B algorithm; analytic image reconstruction algorithms; attenuation correction; data correction; event-by-event correction schemes; image space procedures; iterative image reconstruction algorithms; list-mode data driven reconstruction; positron emission tomography; projection-data based algorithms comparison; rotating 3D PET systems; scatter correction; thorax phantom; Algorithm design and analysis; Attenuation; Bayesian methods; Image analysis; Image reconstruction; Iterative algorithms; Iterative methods; Positron emission tomography; Reconstruction algorithms; Scattering;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium, 1998. Conference Record. 1998 IEEE
  • Conference_Location
    Toronto, Ont.
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-5021-9
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1998.773921
  • Filename
    773921