Firstpage
2060
Abstract
In December 1997 the treatment of patients suffering from head and neck tumours with beams of high-energy 12C ions started at the Gesellschaft fur Schwerionenforschung (GSI) in Darmstadt. By the simultaneous application of positron emission tomography (PET) the treatments are verified in-situ during the irradiation. The PET-based quality assurance procedure is a chain of consecutive steps. The starting point is the three-dimensional treatment planning on the basis of X-ray computed tomograms (CT). During the heavy ion treatment the produced β+-emitter distribution is measured simultaneously. After that the expected temporal and spatial activity distribution is computed. Both images are reconstructed iteratively and filtered. For the analysis the PET and CT images are superimposed
Keywords
positron emission tomography; quality control; radiation therapy; tumours; β+-emitter distribution; C; Gesellschaft fur Schwerionenforschung; PET-based quality assurance procedure; X-ray computed tomograms; filtered images; head/neck tumours; heavy ion radiotherapy; high-energy 12C ions; in-situ treatment verification; iteratively-reconstructed images; Biomedical applications of radiation; Computed tomography; Distributed computing; Head; Medical treatment; Neck; Positron emission tomography; Quality assurance; Tumors; X-ray imaging;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium, 1998. Conference Record. 1998 IEEE
Conference_Location
Toronto, Ont.
ISSN
1082-3654
Print_ISBN
0-7803-5021-9
Type
conf
DOI
10.1109/NSSMIC.1998.773938
Filename
773938
Link To Document