• DocumentCode
    2914598
  • Title

    The RF Voltage Standard Analysis by Parameter Variation

  • Author

    Grno, L.

  • Author_Institution
    Czechoslovak Institute of Metrology
  • fYear
    1988
  • fDate
    7-10 June 1988
  • Firstpage
    375
  • Lastpage
    376
  • Abstract
    The method of the analysis of the thermistor bolometer of the RF voltage standard covering the frequency range up to 1000 MHz is described. The evaluation of the AC/DC difference of the voltage standard is independent of any other RF quantity measurement and serves the accuracy which meets the requirements of the primary standardization.
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements, 1988. CPEM 88 Digest. 1988 Conference on
  • Conference_Location
    Ibaraki, Japan
  • Type

    conf

  • DOI
    10.1109/CPEM.1988.671353
  • Filename
    671353