Abstract
Simultaneous EEG-fMRI studies require an understanding of the noise characteristics of the acquisition environment so that appropriate pre-processing steps may be taken to remove known artifacts from the data stream. Using a phantom approach, we have developed a general methodology for characterizing non-physiologic noise in EEG signal and demonstrate the use of this methodology for a specific MR scanner and EEG data acquisition system configuration. Our results show the δ frequency band is significantly impacted by baseline drift or baseline correction algorithms while the β and γ bands are impacted by residual gradient artifact and gradient corrections.
Keywords
biomedical MRI; electroencephalography; medical signal processing; γ bands; δ frequency band; EEG data acquisition system configuration; EEG signal; baseline correction algorithms; nonphysiologic noise; phantom; residual gradient artifact; simultaneous EEG-fMRI data; specific MR scanner; Cooling; Electrodes; Electroencephalography; Magnetic resonance imaging; Noise; Phantoms; Artifacts; Electric Impedance; Electrodes; Electroencephalography; Humans; Magnetic Resonance Imaging; Nonlinear Dynamics; Phantoms, Imaging; Signal Processing, Computer-Assisted;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society (EMBC), 2010 Annual International Conference of the IEEE
Conference_Location
Buenos Aires
ISSN
1557-170X
Print_ISBN
978-1-4244-4123-5
Type
conf
DOI
10.1109/IEMBS.2010.5626168
Filename
5626168
Link To Document