• DocumentCode
    2924275
  • Title

    A precision 10 mA current shunt is used to calibrate AC current at 1 mA and 100 /spl mu/A levels

  • Author

    Cai, T.-X.

  • Author_Institution
    Simco Electron., Santa Clara, CA, USA
  • fYear
    1998
  • fDate
    6-10 July 1998
  • Firstpage
    31
  • Lastpage
    32
  • Abstract
    The calibration of low AC current at 1 mA and 100 /spl mu/A is described using current divider techniques. Uncertainties are calculated for 9:1 and 99:1 current division ratios. A builddown method is used stepping from 10 mA to 1 mA and from 1 mA to 100 /spl mu/A using a NIST-calibrated 10 mA current shunt as the reference.
  • Keywords
    calibration; electric current measurement; 1 mA to 100 muA; 10 mA; 10 to 1 mA; AC current; NIST; calibration; current division ratio; precision current shunt; Analog-digital conversion; Calibration; Circuits; Consumer electronics; Current measurement; Frequency; Laboratories; NIST; Resistors; Switches;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements Digest, 1998 Conference on
  • Conference_Location
    Washington, DC, USA
  • Print_ISBN
    0-7803-5018-9
  • Type

    conf

  • DOI
    10.1109/CPEM.1998.699744
  • Filename
    699744