DocumentCode
2925586
Title
Calibrating photoreceiver response to 110 GHz
Author
Clement, Tracy S. ; Williams, Dylan F. ; Hale, Paul D. ; Morgan, Juanita M.
Author_Institution
Nat. Inst. of Stand. & Technol., Boulder, CO, USA
Volume
2
fYear
2002
fDate
10-14 Nov. 2002
Firstpage
877
Abstract
We have measured the magnitude and phase responses of a photoreceiver to 110 GHz using a calibrated electro-optic sampling system. The frequency range of the calibration is limited only by our 1 mm coaxial connectors.
Keywords
calibration; coplanar waveguides; electro-optical modulation; equivalent circuits; fibre lasers; laser mode locking; microwave measurement; optical delay lines; optical pulse generation; optical receivers; 1 mm; 110 GHz; calibrated electro-optic sampling system; calibration; coaxial connectors; frequency range; phase responses; photoreceiver response calibration; Coaxial components; Connectors; Coplanar waveguides; Electric variables measurement; Laser theory; Lasers and electrooptics; Optical beams; Optical polarization; Optical waveguides; Sampling methods;
fLanguage
English
Publisher
ieee
Conference_Titel
Lasers and Electro-Optics Society, 2002. LEOS 2002. The 15th Annual Meeting of the IEEE
ISSN
1092-8081
Print_ISBN
0-7803-7500-9
Type
conf
DOI
10.1109/LEOS.2002.1159588
Filename
1159588
Link To Document