Title :
A guarded transfer standard for high resistance measurements
Author_Institution :
Nat. Inst. of Stand. & Technol., Gaithersburg, MD, USA
Abstract :
An improved design for a guarded transfer standard in the resistance range 1 M/spl Omega/ to 100 G/spl Omega/ is described. Existing transfer standards and limitations are reviewed. Interchangeable guard networks are used in the improved transfer standards to ensure complete guarding during all phases of the measurement process thus reducing errors caused by leakages to ground.
Keywords :
electric resistance measurement; leakage currents; measurement errors; resistors; transfer standards; 1 Mohm to 100 Gohm; NIST standards; complete guarding; errors reduction; guard resistors; guarded transfer standard; high resistance measurements; improved design; interchangeable guard networks; leakages to ground; Bridge circuits; Coaxial components; Connectors; Electric resistance; Electrical resistance measurement; Laboratories; Leakage current; Measurement standards; NIST; Resistors;
Conference_Titel :
Precision Electromagnetic Measurements Digest, 1998 Conference on
Conference_Location :
Washington, DC, USA
Print_ISBN :
0-7803-5018-9
DOI :
10.1109/CPEM.1998.699802