DocumentCode
2934130
Title
DC 10:1 Hamon divider with cryogenic contacts
Author
Kyu-Tae Kim ; Sung-Won Kwon
Author_Institution
Korea Res. Inst. of Stand. & Sci., Taejon, South Korea
fYear
1998
fDate
6-10 July 1998
Firstpage
102
Lastpage
103
Abstract
A study on a DC Hamon divider with superconducting contacts (cryo-contacts Hamon divider) is described. The contact resistance error of the switch had been one of the most significant error sources in the conventional Hamon divider. A specially designed cryogenic switch made of Pb was used at liquid helium temperature for a series-parallel connection of the Hamon resistors which form the upper arm of a Wheatstone bridge for a self calibration. The 10:1 ratio uncertainty of the preliminary measurement is expected to easily reach a 0.1 ppm level.
Keywords
bridge circuits; calibration; contact resistance; electric resistance measurement; measurement standards; potentiometers; superconducting switches; voltage measurement; 10:1 ratio uncertainty; DC Hamon divider; Hamon resistors; Wheatstone bridge; contact resistance error; cryo-contacts Hamon divider; cryogenic switch; self calibration; series-parallel connection; superconducting contacts; upper arm; Bridge circuits; Calibration; Contact resistance; Cryogenics; Electrical resistance measurement; Helium; Resistors; Switches; Temperature; Voltage;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements Digest, 1998 Conference on
Conference_Location
Washington, DC, USA
Print_ISBN
0-7803-5018-9
Type
conf
DOI
10.1109/CPEM.1998.699803
Filename
699803