• DocumentCode
    2934144
  • Title

    Optical quantities traceability at NIM-Romania progress toward radiometer based radio-photometry

  • Author

    Simionescu, Mihai ; Seucan, Ainadeu

  • Author_Institution
    Nat. Inst. of Metrol., Romania
  • Volume
    3
  • fYear
    1999
  • fDate
    1999
  • Firstpage
    1869
  • Abstract
    Following the international trend, work aimed at establishing radiometer based spectro-radio-photometric scales was undertaken at NIM Romania. New primary and secondary standards for radiant power, irradiance and spectral transmittance were developed and some results have already been obtained. Standard uncertainties levels so far obtained are of about 0.3% for radiant power and of 0.0005…0.0001 for spectral transmittance (VIS range)
  • Keywords
    calibration; measurement standards; radiometry; NIM-Romania; irradiance; optical quantities traceability; primary standards; radiant power; radiometer based radio-photometry; secondary standards; spectral transmittance; spectro-radio-photometric scales; standard uncertainties levels; Actuators; Bridges; Calibration; Optical receivers; Optical variables control; Photometry; Radiometry; Standards development; Strontium; Transducers;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference, 1999. IMTC/99. Proceedings of the 16th IEEE
  • Conference_Location
    Venice
  • ISSN
    1091-5281
  • Print_ISBN
    0-7803-5276-9
  • Type

    conf

  • DOI
    10.1109/IMTC.1999.776144
  • Filename
    776144