Abstract :
In the traceability transfer process from the national standards of the Istituto Elettrotecnico Nazionale “Galileo Ferraris” (IEN) to standards of calibration laboratories accredited by Servizio di Taratura in Italia (SIT), the use of high precision multifunction instruments, in particular of the digital multimeter, as reference standards turned out to be a good transfer method. In fact it combined the speed and cheapness of transfer to the uncertainty level of the process comparable to that obtained using high precision individual standards, such as 10 V DC voltage standards, standards resistors, AC/DC transfer standards and so on. In the present paper the authors describe how this transfer process is performed and what are the results obtained
Keywords :
calibration; digital multimeters; instruments; stability; transfer standards; DMM; calibration laboratories; digital multimeter; high precision multifunction instruments; national standards; reference standards; stability evaluation; traceability transfer; Automatic voltage control; Calibration; Control systems; Electrical resistance measurement; Electronic equipment testing; Instruments; Laboratories; Measurement standards; Stability; Standards development;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Instrumentation and Measurement Technology Conference, 1999. IMTC/99. Proceedings of the 16th IEEE
Conference_Location :
Venice
ISSN :
1091-5281
Print_ISBN :
0-7803-5276-9
Type :
conf
DOI :
10.1109/IMTC.1999.776145
Filename :
776145
Link To Document :
بازگشت