DocumentCode
2934285
Title
Automatic bridge for comparison of inductances based on difference voltage measurement
Author
Saxena, A.K. ; Saleem, M.
Author_Institution
LF & HF Impedence Stand., New Delhi, India
fYear
1998
fDate
6-10 July 1998
Firstpage
116
Lastpage
117
Abstract
An inductance bridge to compare inductors by 1:1 substitution method is proposed. This bridge is based on difference voltage measurement. Uncertainties of less that 0.01% are possible for inductors from 1 mH to 100 mH at 1 kHz.
Keywords
bridge circuits; calibration; comparators (circuits); computerised instrumentation; inductance measurement; inductors; measurement standards; voltage measurement; 1 kHz; 1:1 substitution method; Maxwell-Wein bridge; PC controlled system; automatic inductance bridge; calibration; difference voltage measurement; inductance standards; inductances comparison; inductors; uncertainties; Bridge circuits; Calibration; Capacitance; Hafnium; Impedance; Inductance; Inductors; Measurement standards; Testing; Voltage measurement;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements Digest, 1998 Conference on
Conference_Location
Washington, DC, USA
Print_ISBN
0-7803-5018-9
Type
conf
DOI
10.1109/CPEM.1998.699810
Filename
699810