Title : 
Calibration of precision measuring amplifiers for strain gauge transducers
         
        
            Author : 
Bohacek, J. ; Nakamura, Y. ; Yoshihiro, K. ; Endo, T.
         
        
            Author_Institution : 
Fac. of Electr. Eng., Czech Tech. Univ., Prague, Czech Republic
         
        
        
        
        
        
            Abstract : 
A method is described that utilizes a ratio device consisting of multi-decade inductive voltage dividers to calibrate precision measuring amplifiers for strain gauge transducers. Most frequently, this device is used to calibrate 2.5 mV/V ranges of the amplifiers, relative standard uncertainty being 5 ppm for the ratio of 0.0025 and at 225 Hz.
         
        
            Keywords : 
bridge circuits; calibration; instrumentation amplifiers; measurement uncertainty; strain gauges; voltage dividers; 225 Hz; calibration; equivalent circuit; multi-decade inductive voltage dividers; precision measuring amplifiers; ratio device; relative standard uncertainty; strain gauge transducers; Admittance; Calibration; Electric variables measurement; Equivalent circuits; Impedance measurement; Laboratories; Measurement standards; Strain measurement; Transducers; Voltage;
         
        
        
        