• DocumentCode
    2935268
  • Title

    Tests of the Recalibration Period of a Drifting Instrument

  • Author

    Liggett, Walter

  • Author_Institution
    National Bureau of Standards, Gaithersburg, MD, USA
  • fYear
    1986
  • fDate
    23-25 Sept. 1986
  • Firstpage
    923
  • Lastpage
    928
  • Abstract
    The use of a drifting instrument requires that an adequately short recalibration period be chosen. After several periods, the calibration data can be used to test the adequacy of the choice. This paper discusses statistical tests of the recalibration period and applies these tests to continuous analyzers for sulfur dioxide. These tests check the sequence of calibrations for evidence of rapid drift-rate changes, which are obscured by random measurement error and by slow drift-rate changes for which the recalibration period is adequate. This paper presents two tests, a test of the second differences of the calibration sequence for normality and a test of the upper part of the spectrum for flatness. Both of these tests are needed because they have different power against different alternatives to constant drift rate. This paper illustrates these tests with two sequences each consisting of about fifty recalibrations of a sulfur dioxide analyzer. Also, the power of the tests and some approximations made in their formulation are investigated by Monte Carlo experiments.
  • Keywords
    Calibration; Curve fitting; Frequency; Instruments; Kalman filters; Monitoring; Robustness; Spline; Testing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    OCEANS '86
  • Conference_Location
    Washington, DC, USA
  • Type

    conf

  • DOI
    10.1109/OCEANS.1986.1160399
  • Filename
    1160399