• DocumentCode
    2936897
  • Title

    Analysis of magnetic field disturbances generated by metallic implants at 0.31T

  • Author

    Guraliuc, Anda R. ; Scartabelli, Vittoria ; Nepa, Paolo ; Manara, Giuliano ; Caramella, Davide

  • Author_Institution
    Dept. of Inf. Eng., Univ. of Pisa, Pisa, Italy
  • fYear
    2011
  • fDate
    3-8 July 2011
  • Firstpage
    124
  • Lastpage
    126
  • Abstract
    Metallic implants placed in a uniform magnetic field of an MRI system generate a magnetic field disturbance. In this paper two paramagnetic material rectangular bars were exposed to a 0.31T static magnetic field of an MRI dedicated system, and the magnetic field disturbance was analyzed. Measurement results were compared with CST EM Studio simulations to confirm the accuracy of the numerical tool when it is used to analyze the electromagnetic disturbance due to the presence of metallic implants in an MRI procedure.
  • Keywords
    biological effects of fields; biomedical MRI; prosthetics; CST EM Studio simulations; MRI system; electromagnetic disturbance; magnetic field disturbances; metallic implants; numerical tool; paramagnetic material rectangular bars; static magnetic field; uniform magnetic field; Bars; Implants; Magnetic field measurement; Magnetic fields; Magnetic flux; Magnetic resonance imaging; Phantoms; 0.31T MRI dedicated system; Magnetic disturbance; Magnetic field interaction; Paramagnetic orthopedic implants;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Antennas and Propagation (APSURSI), 2011 IEEE International Symposium on
  • Conference_Location
    Spokane, WA
  • ISSN
    1522-3965
  • Print_ISBN
    978-1-4244-9562-7
  • Type

    conf

  • DOI
    10.1109/APS.2011.5996656
  • Filename
    5996656