• DocumentCode
    2937664
  • Title

    A small size dipole E-field sensor for calibration of field distribution in EMP simulator

  • Author

    Li-Hua Shi ; Rong-ren Si ; Qing Si ; Yan-xin Li

  • Author_Institution
    Nat. Key Lab. on Electromagn. Environ. Effects & Electro-Opt. Eng., PLA Univ. of Sci. & Technol., Nanjing, China
  • fYear
    2012
  • fDate
    6-9 Nov. 2012
  • Firstpage
    251
  • Lastpage
    254
  • Abstract
    The calibration of field distribution in EMP simulator is discussed. A small size dipole sensor with built-in optic transmitter and fiber optic output connection is designed to measure field distribution in various points of the EMP simulator. Influence of the dipole sensor to the field distribution is first analyzed by numerical simulation and the actual field distribution is measured by the dipole. A correction coefficient is proposed based on the measurement results. The available working space in EMP simulator for sensor calibration is also defined.
  • Keywords
    calibration; dipole antennas; electric field measurement; electromagnetic pulse; fibre optic sensors; numerical analysis; EMP simulator; built-in optic transmitter; correction coefficient; dipole antenna; fiber optic output connection; field distribution calibration; numerical simulation; small size dipole E-field sensor; Manganese; Optical fiber sensors; Optical fibers; Probes; EMP; calibration; dipole antenna; sensor; standard field;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Environmental Electromagnetics (CEEM), 2012 6th Asia-Pacific Conference on
  • Conference_Location
    Shanghai
  • Print_ISBN
    978-1-4673-0030-8
  • Type

    conf

  • DOI
    10.1109/CEEM.2012.6410614
  • Filename
    6410614