Improved resolution via 3D iterative reconstruction for PET volume imaging
         
        
            Author : 
Liow, J.-S. ; Strother, S.C. ; Rottenberg, D.A.
         
        
            Author_Institution : 
Minnesota Univ., Minneapolis, MN, USA
         
        
        
        
            fDate : 
30 Oct-5 Nov 1994
         
        
        
            Abstract : 
The authors have implemented iterative filtered backprojection (IFBP) and maximum likelihood by expectation maximization (ML-EM) algorithms in 3D space and applied them to phantom and real PET data. Transaxial resolution improves ≈50% and axial resolution improves ≈15% for IFBP at 15 iterations without a sieve compared to FBP. With a sieve, the improvements are reduced to ≈6%. 3D ML-EM reconstruction shows similar resolution improvement with a much slower convergence rate compared to IFBP. The improvements in resolution from both IFBP and ML-EM are apparent in 3D FDG brain data
         
        
            Keywords : 
image reconstruction; iterative methods; medical image processing; positron emission tomography; 3D FDG brain data; 3D iterative reconstruction; 3D space; PET volume imaging; axial resolution; convergence rate; image resolution improvement; iterative filtered backprojection algorithm; maximum likelihood by expectation maximization algorithm; medical diagnostic imaging; nuclear medicine; sieve; transaxial resolution; Apertures; Attenuation; Cameras; Filters; Image reconstruction; Image resolution; Positron emission tomography; Reconstruction algorithms; Smoothing methods; Spatial resolution;
         
        
        
        