Abstract :
This paper describes two useful metrics to estimate the ligament imbalance for post Total Knee Arthroplasty (TKA) scenario: the center of pressure and the net moments (varus-valgus and anterior posterior). Both metrics have been evaluated using high level models and experimental measurement. Self-powered analog and digital architectures for the center of pressure are elaborated here and complies with the Medical Implant Communication Service (MICS) standard. It is shown that the analog architecture is advantageous in terms of surface area and overall power consumption.
Keywords :
biomedical measurement; prosthetic power supplies; Medical Implant Communication Service standard; autonomous measurement system; digital architectures; ligament imbalance metrics; overall power consumption; post TKA; post total knee arthroplasty scenario; self-powered analog architectures; surface area; Computer architecture; Force; Knee; Ligaments; Measurement; Power demand; Prosthetics; Arthroplasty, Replacement, Knee; Humans; Ligaments, Articular; Telemetry; Tibia;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society (EMBC), 2010 Annual International Conference of the IEEE
Conference_Location :
Buenos Aires
ISSN :
1557-170X
Print_ISBN :
978-1-4244-4123-5
Type :
conf
DOI :
10.1109/IEMBS.2010.5627322
Filename :
5627322
Link To Document :
بازگشت