DocumentCode
2943323
Title
Compensation of the fluid temperature variation in a hot-wire anemometer
Author
Oliveira, Amauri ; Freire, Raimundo C S ; Deep, Gurdip S.
Author_Institution
EP-DEE, UFBA, Brazil
Volume
2
fYear
1997
fDate
19-21 May 1997
Firstpage
1377
Abstract
In conventional constant-temperature hot-wire anemometers, the effect of fluid temperature variation on the measurement is compensated by using an additional sensor sensitive to the fluid temperature, but almost insensitive to fluid velocity. In this paper, a measurement configuration based on two similar heated sensors maintained at two different temperatures, interacting identically with the measurand is proposed and formulated. The possibility of using only one sensor to measure fluid velocity, independent of the fluid temperature is also proposed
Keywords
anemometers; bridge instruments; circuit feedback; error compensation; pulse width modulation; signal processing; thermistors; compensation; constant-temperature hot-wire anemometers; feedback; fluid temperature variation; fluid velocity; heated sensors; hot-wire anemometer; pulse width modulation; Fluid flow; Fluid flow measurement; Negative feedback; Sensor phenomena and characterization; Temperature dependence; Temperature sensors; Thermal sensors; Thermoresistivity; Velocity measurement; Voltage;
fLanguage
English
Publisher
ieee
Conference_Titel
Instrumentation and Measurement Technology Conference, 1997. IMTC/97. Proceedings. Sensing, Processing, Networking., IEEE
Conference_Location
Ottawa, Ont.
ISSN
1091-5281
Print_ISBN
0-7803-3747-6
Type
conf
DOI
10.1109/IMTC.1997.612424
Filename
612424
Link To Document