Title :
Calibration of dissipation factor standards
Author :
Simmon, E.D. ; FitzPatrick, G.J. ; Petersons, O.
Author_Institution :
Nat. Inst. of Stand. & Technol., Gaithersburg, MD, USA
Abstract :
Dissipation factor (DF) standards obtained by connecting a shielded three-terminal capacitor in series with a shielded resistor have been developed for calibration purposes. An analysis of these DF standards, including precautions in their construction and use, is presented and calibration procedures using the NIST high voltage capacitance bridge is discussed.
Keywords :
bridge instruments; calibration; capacitors; dielectric loss measurement; measurement standards; NIST high voltage capacitance bridge; calibration; dissipation factor standard; shielded resistor; shielded three terminal capacitor; Admittance; Calibration; Capacitance; Capacitors; Connectors; Joining processes; Measurement standards; NIST; Resistors; Voltage;
Conference_Titel :
Precision Electromagnetic Measurements Digest, 1998 Conference on
Conference_Location :
Washington, DC, USA
Print_ISBN :
0-7803-5018-9
DOI :
10.1109/CPEM.1998.699867