Validation of a transit time blood flow meter used for coronary bypass surgery
Author :
Ortega-Palacios, R. ; Moreno, E. ; Leija, L. ; Vera, A.
Author_Institution :
Electr. Eng. Dept., CINVESTAV-IPN, Mexico City, Mexico
fYear :
2010
fDate :
Aug. 31 2010-Sept. 4 2010
Firstpage :
178
Lastpage :
181
Abstract :
The validation of a blood flow meter is presented. This flow meter is based on the transit time method that uses two ultrasound transducers and a reflector. The measuring is based on the time difference of the upstream and the downstream ultrasound transit time. The validation was based on the norm ANSI Std. N42.17A-2003 of blood flow protocol. We used a flow simulator system with 3 mm pipe, bi-distilled and desgasificated water as fluid. Flow simulator system uses a constant flow pump, regulated by a valve, and a peristaltic flow pump, regulated by a variable voltage source, like flow source through the pipe.
Keywords :
blood flow measurement; peristaltic flow; surgery; transit time devices; constant flow pump; coronary bypass surgery; peristaltic flow pump; reflector; transit time blood flow meter; ultrasound transducers; valve; variable voltage source; Acoustics; Blood; Transducers; Ultrasonic imaging; Ultrasonic variables measurement; Valves; Voltage control; Blood Flow Velocity; Coronary Artery Bypass; Humans; Least-Squares Analysis; Models, Cardiovascular; Reproducibility of Results; Rheology; Signal Processing, Computer-Assisted; Ultrasonography;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society (EMBC), 2010 Annual International Conference of the IEEE
Conference_Location :
Buenos Aires
ISSN :
1557-170X
Print_ISBN :
978-1-4244-4123-5
Type :
conf
DOI :
10.1109/IEMBS.2010.5627483
Filename :
5627483
Link To Document :
بازگشت