DocumentCode
2948662
Title
RUPERT closed loop control design
Author
Zhang, Hang ; Balasubramanian, Sivakumar ; Wei, Ruihua ; Austin, Hiroko ; Buchanan, Sharon ; Herman, Richard ; He, Jiping
Author_Institution
Sch. of Biol. & Health Syst. Eng., Arizona State Univ., Tempe, AZ, USA
fYear
2010
fDate
Aug. 31 2010-Sept. 4 2010
Firstpage
3686
Lastpage
3689
Abstract
Robot-assisted rehabilitation is an active area of research in the field of stroke rehabilitation. RUPERT is a wearable robotic exoskeleton powered by pneumatic muscle actuators. In this study, we described the structure of the controllers for the five degrees of freedom currently used by RUPERT. We applied the RUPERT on 6 stroke patients to provide robot-assisted rehabilitation therapy in a clinical study. Statistical χ2 test on the proportion of successfully reaching targets showed that 3 out of the 6 patients demonstrated significant improvement in reaching targets successfully, and the remaining 3 did not show performance improvement or deterioration. We plan to implement the RUPERT in the patient´s house for easier access and more frequent use. More significant performance results are expected.
Keywords
closed loop systems; diseases; medical robotics; patient rehabilitation; pneumatic actuators; RUPERT closed loop control design; controller structure; pneumatic muscle actuators; robot assisted rehabilitation therapy; stroke rehabilitation; wearable robotic exoskeleton; Generators; Joints; Medical treatment; Robot kinematics; Testing; Training; χ2 test; PID controller; feedback control; rehabilitation robot; stroke; Arm; Biofeedback, Psychology; Computer-Aided Design; Humans; Movement Disorders; Orthotic Devices; Prosthesis Design; Robotics;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society (EMBC), 2010 Annual International Conference of the IEEE
Conference_Location
Buenos Aires
ISSN
1557-170X
Print_ISBN
978-1-4244-4123-5
Type
conf
DOI
10.1109/IEMBS.2010.5627647
Filename
5627647
Link To Document