Catheter flow sensor system and breathing measurements in rabbit
Author
Matsuyama, Takashi ; Yamazaki, Yasuyuki ; Shikano, T. ; Shikida, Mitsuhiro ; Matsushima, Miyoko ; Kawabe, Takashi
fYear
2012
fDate
28-31 Oct. 2012
Firstpage
1
Lastpage
4
Abstract
A catheter flow sensor system is developed to directly measure the breathing characteristics in small bronchi. The flow sensor is fabricated on a 15-μm thick biocompatible parylene HT film. The film is mounted inside the surface of a thermal shrinkable tube. Finally, a handling tube is connected to one end of the fabricated sensor. A guide tube and a fiberscope are used to insert the sensor into a rabbit bronchus. The fore-end position of the guide tube is confirmed from images taken by the fiberscope. The sensor directly evaluates the breathing characteristics of rabbits having and not having a pneumothorax. The per-cycle flow volume for a rabbit having a pneumothorax is found to be only 70-77% of that for a rabbit not having one.
Keywords
catheters; flow sensors; biocompatible parylene HT film; breathing measurement; catheter flow sensor system; fiberscope; guide tube; rabbit bronchus; size 15 mum; small bronchi; thermal shrinkable tube; Catheters; Electron tubes; Films; Fluid flow measurement; Optical fiber sensors; Rabbits; Respiratory system;