Title :
Test system for clinical force platforms
Author :
Sanhueza, Germán Blanchard
Author_Institution :
Sci. Fac., Univ. de Valparaiso, Valparaiso, Chile
fDate :
Aug. 31 2010-Sept. 4 2010
Abstract :
Our team has built a system for testing and calibration of clinical Force Platforms, able to generate a circular path Center of Pressure (CoP) that can be accurately applied in the horizontal plane at a given travel speed, selectable radius and vertical force. The CoP generated, which is applied directly onto the platform, is used as a precise reference for both static and dynamic measurements, to check the operation of unknown Force Platforms performance characteristics or to verify specifications indicated by the Force Platform´s manufacturer. To confirm the feasibility and effectiveness of the proposed test apparatus, it was applied to a commercial clinical usage Force Platform manufactured by artOficio LTDA. The main advantages of the proposed Test System are the accuracy in fixing or setting a direct reference and the ability to maintain continuous monitoring through the generation of flexible and stable circular patterns during the tests.
Keywords :
biomechanics; biomedical measurement; calibration; force; testing; circular path center of pressure; clinical force platforms; Accuracy; Dynamics; Force; Force measurement; Frequency measurement; Frequency modulation; Position measurement; Biomechanics; Fourier Analysis; Humans; Physical Medicine; Postural Balance; Pressure;