• DocumentCode
    2954411
  • Title

    A three-way, on-site comparison of the 10 V-Josephson voltage standards of the PTB, the SP and the BIPM

  • Author

    Reymann, D. ; Witt, T.J. ; Eklund, G. ; Pajander, H. ; Nilsson, H. ; Behr, R. ; Funck, T. ; Muller, F.

  • Author_Institution
    Bur. Int. des Poids et Mesures, Sevres, France
  • fYear
    1998
  • fDate
    6-10 July 1998
  • Firstpage
    351
  • Lastpage
    352
  • Abstract
    Comparisons of the 10 V Josephson array voltage standard (JAVS) of the Bureau International des Poids et Mesures (BIPM) were made at the Physikalisch-Technische Bundesansalt (PTB), Braunschweig, Germany with those of the PTB and of the Swedish National Testing and Research Institute (SP), Boras, Sweden, in January 1998. Although the results agree within an overall uncertainty of less than 1 part in 10/sup 9/, a small, but interesting, dependence on microwave power was observed in the DC output of one array.
  • Keywords
    measurement standards; superconducting junction devices; voltage measurement; 10 V; BIPM; Josephson array voltage standard; PTB; SP; Bismuth; Measurement standards; Metrology; Microwave antenna arrays; NIST; Sensor arrays; Testing; Uncertainty; Video recording; Voltage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements Digest, 1998 Conference on
  • Conference_Location
    Washington, DC, USA
  • Print_ISBN
    0-7803-5018-9
  • Type

    conf

  • DOI
    10.1109/CPEM.1998.699946
  • Filename
    699946