  • Abstract
    Two Josephson Array Voltage Standard Systems using different calibration algorithms are operated at the National Institute of Standards and Technology (NIST). A manual switch system is designed to make system comparisons by measuring against the 1.018 V of three Zener references. The difference of the average of the three Zener references is smaller than 1 nV.
  • Keywords
    Zener diodes; calibration; measurement standards; superconducting junction devices; voltage measurement; 1.018 V; Josephson array voltage standard; NIST; Zener reference; calibration algorithm; manual switch; Calibration; Frequency; Hardware; Instruments; Josephson junctions; NIST; Performance evaluation; Physics; Switches; Voltage;