DocumentCode
2959035
Title
CMOS BDJ detector array with charge preamplifiers for sensitive biochemical analysis
Author
Carrillo, G. ; Lu, G.-N. ; Pittet, P. ; Zhao, K.
Author_Institution
Univ. Claude Bernard Lyon 1, Villeurbanne
fYear
2006
fDate
10-13 Dec. 2006
Firstpage
367
Lastpage
370
Abstract
For instrumental implementation of biochemical microanalysis techniques, we have designed a CMOS BDJ (buried double p-n junction) detector array and its associated charge amplifiers. The adopted amplifier configuration allows a high-sensitivity performance (100 V/lx.s@555 nm). The detector chip can operate in low-bias and low-temperature conditions, so as to lower the dark currents of the detector and to increase the integration time of the charge amplifiers. In such operating conditions, both MDSmin (minimum detectable signal) and DR (dynamic range) can be significantly improved. We have obtained, e.g., at -10degC, MDSmin = 1mulx@555 nm and DR = 135 dB.
Keywords
CMOS integrated circuits; biochemistry; biological techniques; chemical analysis; lab-on-a-chip; preamplifiers; CMOS BDJ detector array; biochemical microanalysis techniques; buried double p-n junction detector array; charge preamplifiers; detector chip; sensitive biochemical analysis; Biochemical analysis; Detectors; Dynamic range; Instruments; Optical amplifiers; P-n junctions; Photodetectors; Photodiodes; Preamplifiers; Sensor arrays;
fLanguage
English
Publisher
ieee
Conference_Titel
Electronics, Circuits and Systems, 2006. ICECS '06. 13th IEEE International Conference on
Conference_Location
Nice
Print_ISBN
1-4244-0395-2
Electronic_ISBN
1-4244-0395-2
Type
conf
DOI
10.1109/ICECS.2006.379801
Filename
4263379
Link To Document