Title : 
Control variables and algorithms for in series cardiac assistance
         
        
            Author : 
Cui, T. ; Li, J.K.-J. ; Welkowitz, W. ; Petrucelli, S.
         
        
            Author_Institution : 
Dept. of Biomed. Eng., Rutgers Univ., Piscataway, NJ, USA
         
        
        
        
        
        
            Abstract : 
Hemodynamic control variables such as left atrial pressure, heart rate, and tension time index have been examined for in-series cardiac assistance using the intraaortic balloon pump. Automatic control algorithms have been constructed based on analog models of the assisted circulation. These are reviewed and a scheme for implantable automatic controlled cardiac assist devices are proposed
         
        
            Keywords : 
artificial organs; biocontrol; cardiology; haemodynamics; analog models; haemodynamic control variables; heart rate; in series cardiac assistance; intraaortic balloon pump; left atrial pressure; tension time index; Automatic control; Biomedical measurements; Blood flow; Electrocardiography; Heart rate; Hemodynamics; Optimal control; Performance analysis; Pressure measurement; Timing;
         
        
        
        
            Conference_Titel : 
Bioengineering Conference, 1988., Proceedings of the 1988 Fourteenth Annual Northeast
         
        
            Conference_Location : 
Durham, NH
         
        
        
            DOI : 
10.1109/NEBC.1988.19328