Asia/Pacific Metrology Program (APMP) comparison of DC voltage-1996/97
Author
Liu, L.X. ; Sim, T.Y. ; Tan, V.K.S. ; Chua, H.A.
Author_Institution
Nat. Meas. Centre, Singapore
fYear
1998
fDate
6-10 July 1998
Firstpage
562
Lastpage
563
Abstract
An APMP international comparison of DC voltage, with 15 participants, using a Zener cell as traveling standard was carried out from February 1996 to December 1997. The final results show an agreement of the majority of participating laboratories of better than 0.2 ppm for 1.018 V output.
Keywords
Zener diodes; calibration; transfer standards; voltage measurement; 1.018 V; Asia/Pacific Metrology Program comparison; DC voltage; Zener cell; calibration; international comparison; transfer standard; traveling standard; Asia; Collision mitigation; Humidity; Laboratories; Measurement standards; Metrology; Particle measurements; Productivity; Standards Board; Voltage;