Title :
Asia/Pacific Metrology Program (APMP) comparison of DC voltage-1996/97
Author :
Liu, L.X. ; Sim, T.Y. ; Tan, V.K.S. ; Chua, H.A.
Author_Institution :
Nat. Meas. Centre, Singapore
Abstract :
An APMP international comparison of DC voltage, with 15 participants, using a Zener cell as traveling standard was carried out from February 1996 to December 1997. The final results show an agreement of the majority of participating laboratories of better than 0.2 ppm for 1.018 V output.
Keywords :
Zener diodes; calibration; transfer standards; voltage measurement; 1.018 V; Asia/Pacific Metrology Program comparison; DC voltage; Zener cell; calibration; international comparison; transfer standard; traveling standard; Asia; Collision mitigation; Humidity; Laboratories; Measurement standards; Metrology; Particle measurements; Productivity; Standards Board; Voltage;
Conference_Titel :
Precision Electromagnetic Measurements Digest, 1998 Conference on
Conference_Location :
Washington, DC, USA
Print_ISBN :
0-7803-5018-9
DOI :
10.1109/CPEM.1998.700056