• DocumentCode
    2970192
  • Title

    Improvement of the DC-voltage accuracy in production-line

  • Author

    Yoshida, H. ; Kozakai, T.

  • Author_Institution
    Advantest Corp., Tokyo, Japan
  • fYear
    1998
  • fDate
    6-10 July 1998
  • Firstpage
    570
  • Lastpage
    571
  • Abstract
    In order to reflect the accuracy of the production-line, the precise, but noisy DC voltage output of the PWM (Pulse Width Modulation) type calibrator is successfully measured by the Josephson instrument combined with a newly developed filter.
  • Keywords
    calibration; low-pass filters; passive filters; production testing; pulse width modulation; superconducting junction devices; voltage measurement; DC voltage accuracy; JJAVS; Josephson instrument; Josephson junction array voltage standard; PWM type calibrator; filter; production-line measurement; pulse width modulation; Frequency; Instruments; Low pass filters; Measurement standards; Noise measurement; Pulse measurements; Pulse width modulation; Space vector pulse width modulation; Time domain analysis; Voltage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements Digest, 1998 Conference on
  • Conference_Location
    Washington, DC, USA
  • Print_ISBN
    0-7803-5018-9
  • Type

    conf

  • DOI
    10.1109/CPEM.1998.700060
  • Filename
    700060