Magnetometer calibration set-up controlled by nuclear magnetic resonance
Author :
Weyand, K.
Author_Institution :
Phys. Tech. Bundesanstalt, Berlin, Germany
fYear :
1998
fDate :
6-10 July 1998
Firstpage :
580
Lastpage :
581
Abstract :
In order to calibrate magnetometer sensors in the field range from 2 mT to 150 mT a measuring setup consisting of a water-cooled multilayer solenoid and an NMR-magnetometer of the marginal oscillator type has been developed. With a coil constant of about 20 mT/A the field coil provides a field profile with two flat regions, where the field maxima are equal within 10/sup -5/ and with a distance of more than 40 mm so that both magnetometer readings can simultaneously be compared. The NMR-magnetometer operates with samples having an absorption line width of 2 /spl mu/T with a sample volume of 2 cm/sup 3/ for fields below 20 mT and of 1 cm/sup 3/ for higher fields.
Keywords :
calibration; magnetometers; nuclear magnetic resonance; solenoids; 2 to 150 mT; absorption line width; calibration set-up; coil constant; field coil; field maxima; field profile; flat regions; magnetometer calibration; magnetometer readings; marginal oscillator type; nuclear magnetic resonance; sample volume; water-cooled multilayer solenoid; Absorption; Calibration; Coils; Magnetic field measurement; Magnetic multilayers; Magnetic sensors; Magnetometers; Nuclear magnetic resonance; Oscillators; Packaging;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements Digest, 1998 Conference on
Conference_Location :
Washington, DC, USA
Print_ISBN :
0-7803-5018-9
Type :
conf
DOI :
10.1109/CPEM.1998.700065
Filename :
700065
Link To Document :
بازگشت