DocumentCode :
2972527
Title :
A rapid quantitative measurement using the electro-microchip for ketamine and luteinizing hormone detection
Author :
Yeh, Chia-Hsien ; Wang, Wei-Ting ; Sun, Zheng-Kai ; Lin, Yu-Cheng
Author_Institution :
Eng. Sci., Nat. Cheng Kung Univ., Tainan, Taiwan
fYear :
2011
fDate :
28-31 Oct. 2011
Firstpage :
940
Lastpage :
942
Abstract :
This study successfully developed the electro-microchip to rapidly detect the ketamine (KET) and luteinizing hormone (LH) concentration by using the competitive and sandwich immunoassay, respectively. Our strategy was that the AuNPs were introduced into the electro-microchip by the specific binding of the antibodies and constructs a “bridge” between two electrodes allowing the electrons to pass. The detectable concentration limits for KET and LH were 50 ng/mL and 20 mIU/mL, respectively. Therefore, the developed electro-microchip has advantages for the smaller sample volume (30 μL), rapid quantitative measurement, and smaller detectable concentration than the commercial detection products (KET: 100 ng/mL, and LH: 80 mIU/mL).
Keywords :
biosensors; chemical analysis; chemical variables measurement; microelectrodes; antibodies; electro-microchip; immunoassay; ketamine; luteinizing hormone detection; rapid quantitative measurement; specific binding; Electrodes; Glass; Gold; Immune system; Impedance; Silicon; Surface impedance;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Sensors, 2011 IEEE
Conference_Location :
Limerick
ISSN :
1930-0395
Print_ISBN :
978-1-4244-9290-9
Type :
conf
DOI :
10.1109/ICSENS.2011.6127293
Filename :
6127293
Link To Document :
بازگشت