• DocumentCode
    2976287
  • Title

    Reciprocity calibration of hydrophones in the Megahertz frequency range [biomedical ultrasound applications]

  • Author

    Labuda, Cecille ; Raymond, Jason L. ; Church, Charles C.

  • Author_Institution
    Nat. Center for Phys. Acoust., Mississippi Univ., MS, USA
  • Volume
    3
  • fYear
    2004
  • fDate
    23-27 Aug. 2004
  • Firstpage
    1595
  • Abstract
    In biomedical ultrasound, experimental work may be performed at various temperatures whereas hydrophone calibration data are typically specified at only one temperature. It is necessary to know the temperature dependence, if any, of hydrophone sensitivity. Comparison of the results of a three-transducer reciprocity calibration of an Imotec type 80-0.5-4.0 passive PVDF probe at 3°C, 15°C and 37°C to the calibration at 22°C show that within the limits of precision of this experiment, there is no difference in hydrophone sensitivity at these four temperatures.
  • Keywords
    biomedical ultrasonics; calibration; hydrophones; 15 degC; 22 degC; 3 degC; 37 degC; biomedical ultrasound; hydrophone reciprocity calibration; hydrophone sensitivity temperature dependence; passive PVDF probe; three-transducer reciprocity calibration; Blood; Calibration; Frequency; Probes; Samarium; Sonar equipment; Temperature dependence; Temperature distribution; Temperature sensors; Ultrasonic imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 2004 IEEE
  • ISSN
    1051-0117
  • Print_ISBN
    0-7803-8412-1
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2004.1418125
  • Filename
    1418125