DocumentCode
2976287
Title
Reciprocity calibration of hydrophones in the Megahertz frequency range [biomedical ultrasound applications]
Author
Labuda, Cecille ; Raymond, Jason L. ; Church, Charles C.
Author_Institution
Nat. Center for Phys. Acoust., Mississippi Univ., MS, USA
Volume
3
fYear
2004
fDate
23-27 Aug. 2004
Firstpage
1595
Abstract
In biomedical ultrasound, experimental work may be performed at various temperatures whereas hydrophone calibration data are typically specified at only one temperature. It is necessary to know the temperature dependence, if any, of hydrophone sensitivity. Comparison of the results of a three-transducer reciprocity calibration of an Imotec type 80-0.5-4.0 passive PVDF probe at 3°C, 15°C and 37°C to the calibration at 22°C show that within the limits of precision of this experiment, there is no difference in hydrophone sensitivity at these four temperatures.
Keywords
biomedical ultrasonics; calibration; hydrophones; 15 degC; 22 degC; 3 degC; 37 degC; biomedical ultrasound; hydrophone reciprocity calibration; hydrophone sensitivity temperature dependence; passive PVDF probe; three-transducer reciprocity calibration; Blood; Calibration; Frequency; Probes; Samarium; Sonar equipment; Temperature dependence; Temperature distribution; Temperature sensors; Ultrasonic imaging;
fLanguage
English
Publisher
ieee
Conference_Titel
Ultrasonics Symposium, 2004 IEEE
ISSN
1051-0117
Print_ISBN
0-7803-8412-1
Type
conf
DOI
10.1109/ULTSYM.2004.1418125
Filename
1418125
Link To Document