• DocumentCode
    2977258
  • Title

    Evaluation of 100 ampere, 100 kHz transconductance amplifiers

  • Author

    Hess, D.T. ; Clarke, K.K.

  • Author_Institution
    Clarke-Hess Commun. Res. Corp., New York, NY, USA
  • fYear
    1998
  • fDate
    6-10 July 1998
  • Firstpage
    611
  • Lastpage
    612
  • Abstract
    This paper presents experimental evaluation of the performance of more than ten production units of a 100 A, 100 kHz transconductance amplifier. In addition to the stability, uncertainty, distortion and frequency response, these results indicate that the output impedance may be closely modeled by a parallel combination of a resistor and a capacitor. This model then proves to be very useful for correcting the errors introduced into the high-frequency high current measurements by the parasitic load inductance.
  • Keywords
    calibration; electric current measurement; frequency response; harmonic distortion; measurement errors; measurement standards; measurement uncertainty; operational amplifiers; 100 A; 100 kHz; amplifier performance; automated range setting; calibration; distortion; error correction; frequency response; high-frequency high current measurements; linearity; output impedance; parallel resistor-capacitor combination; parasitic load inductance; stability; transconductance amplifiers; uncertainty; Capacitors; Current measurement; Distortion measurement; Error correction; Frequency response; Impedance; Stability; Transconductance; Uncertainty; Voltage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements Digest, 1998 Conference on
  • Conference_Location
    Washington, DC, USA
  • Print_ISBN
    0-7803-5018-9
  • Type

    conf

  • DOI
    10.1109/CPEM.1998.700105
  • Filename
    700105