    An accurate RMS/DC voltage comparator

  • Author

    Arseneau, R. ; Zelle, J.

  • Author_Institution
    Nat. Res. Council of Canada, Ottawa, Ont., Canada
  • fYear
    1990
  • fDate
    13-15 Feb 1990
  • Firstpage
    19
  • Abstract
    The authors describe a new instrument which is designed to operate as a null meter with a range of ±1000 p.p.m. of typical test voltages used in high-accuracy AC calibrations. It is based on a commercially available dual-heater-type converter in a standard integrated circuit package. The accuracy of the instrument is better than 100 p.p.m. for frequencies of 50 Hz to 200 kHz. It can, however, be used for frequencies up to 1 MHz at reduced accuracy
  • Keywords
    calibration; comparators (circuits); voltage measurement; 50 Hz to 1 MHz; AC calibrations; AC voltage measurement; RMS/DC voltage comparator; dual-heater-type converter; null meter; standard integrated circuit package; test voltages; Calibration; Counting circuits; Frequency; Hydrogen; Instruments; Integrated circuit measurements; Power measurement; Switches; Testing; Voltage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference, 1990. IMTC-90. Conference Record., 7th IEEE
  • Conference_Location
    San Jose, CA
  • Type

    conf

  • DOI
    10.1109/IMTC.1990.65950
  • Filename