DocumentCode
2994926
Title
Measurements of Tissue Temperatures during Semiconductor Laser Welding Process
Author
Zajac, A. ; Polakowski, Henryk ; Piatkowski, Tadeusz ; Kastek, Mariusz
Author_Institution
Inst. of Optoelectron., Mil. Univ. of Technol., Warsaw, Poland
fYear
2012
fDate
21-23 May 2012
Firstpage
1
Lastpage
4
Abstract
During the research the effect of coagulation of biological tissues was investigated, caused by pulse SSL and CW semiconductor laser illumination (λ = 1000 nm). The state and temperature of soft biological tissues exposed to laser radiation was observed in real time, using thermographic and optical methods. The temperature changes, including maximal values, were evaluated by analyzing recorded thermal image. The measured scattered radiation of diagnostic laser (λ = 635 nm) was used to assess the tissue condition during the laser welding process. Exemplary results of the laboratory research on these methods and an interpretation of the results are presented.
Keywords
biological effects of laser radiation; biological tissues; biomedical measurement; biothermics; radiation therapy; semiconductor lasers; temperature measurement; coagulation effects; diagnostic laser scattered radiation; laser radiation exposure; optical methods; pulsed CW semiconductor laser illumination; pulsed SSL semiconductor laser illumination; semiconductor laser welding process; soft biological tissues; thermographic methods; tissue temperature changes; tissue temperature measurements; wavelength 1000 nm; wavelength 635 nm; Coagulation; Laser beams; Laser modes; Measurement by laser beam; Semiconductor lasers; Temperature measurement; Welding;
fLanguage
English
Publisher
ieee
Conference_Titel
Photonics and Optoelectronics (SOPO), 2012 Symposium on
Conference_Location
Shanghai
ISSN
2156-8464
Print_ISBN
978-1-4577-0909-8
Type
conf
DOI
10.1109/SOPO.2012.6270546
Filename
6270546