DocumentCode :
3002485
Title :
Interference reduction in nerve cuff electrode recordings-a new approach
Author :
Rahal, Mouhamed ; Winter, Jeff ; Taylor, John ; Donaldson, Nick
Author_Institution :
Dept. of Electr. Eng., Univ. Coll. London, UK
Volume :
6
fYear :
1999
fDate :
36342
Firstpage :
493
Abstract :
A method is presented to reduce the interference pickup in nerve cuff recordings. The gains of differential amplifiers connected to the true tripole nerve cuff arrangement are tuned adaptively to null the residual EMG. Simulation results show that extraction of the neural signal is possible using this method without the need for high order filtering
Keywords :
biomedical electrodes; biomedical electronics; circuit tuning; electromyography; instrumentation amplifiers; interference suppression; low-power electronics; neuromuscular stimulation; adaptively tuned amplifiers; differential amplifiers; functional electrical stimulation; interference pickup; interference reduction; nerve cuff electrode recordings; neural signal extraction; residual EMG nulling; tetraplegic patients; true tripole nerve cuff arrangement; Biomedical electrodes; Biomedical engineering; Biomembranes; Differential amplifiers; Electromyography; Impedance; Insulation; Interference; Muscles; Physics;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Circuits and Systems, 1999. ISCAS '99. Proceedings of the 1999 IEEE International Symposium on
Conference_Location :
Orlando, FL
Print_ISBN :
0-7803-5471-0
Type :
conf
DOI :
10.1109/ISCAS.1999.780202
Filename :
780202
Link To Document :
بازگشت