  • Abstract
    High frequency ultrasonic transducers were fabricated using lead indium niobate-lead magnesium niobate-lead titanate (0.24PIN-0.44PMN-0.32PT) as the active piezoelectric material. The measured center frequency and bandwidth of the device were 35 MHz and 48 % respectively. The electrical impedance magnitude was 71 Omega and phase was - 38deg. Insertion loss was measured to be 15 dB. A phantom which consists of 20-mum tungsten wires was imaged to assess spatial resolution of the transducer.
  • Keywords
    biomedical ultrasonics; phantoms; piezoelectric transducers; ultrasonic transducers; PIN-PMN-PT single crystal high frequency ultrasound transducer; electrical impedance magnitude; insertion loss; lead indium niobate-lead magnesium niobate-lead titanate; medical application; phantom; piezoelectric material; spatial resolution; Biomedical equipment; Biomedical transducers; Frequency; Indium; Magnesium; Medical services; Piezoelectric transducers; Ultrasonic imaging; Ultrasonic transducers; Ultrasonic variables measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 2008. IUS 2008. IEEE
  • Conference_Location
    Beijing
  • Print_ISBN
    978-1-4244-2428-3
  • Electronic_ISBN
    978-1-4244-2480-1
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2008.0348
  • Filename
    4803566