Title :
A comprehensive error model for polarimetric measurement systems
Author :
Bouzari, H. ; Tavaksli, A. ; Heydar, H.
Author_Institution :
Amirkabir Univ. of Technol., Tehran, Iran
Abstract :
In polarimetric measurement systems, the calibration procedure is based on an error model. A system consisting of a conventional automatic vector network analyzer test set, an OMT polariser, a horn antenna, and an anechoic chamber is calibrated with a 30 term error model. We present the 18 term model used for the combination of the OMT, the horn and the chamber. The model overcomes the impedance mismatches, non-symmetrical radiation patterns in horizontal and vertical planes, leakage, and wave reflections. A simple and precise three-stage calibration procedure is proposed that considers all the 18 error terms. Theoretical results are in very good agreement with experimental results.
Keywords :
S-parameters; anechoic chambers; antenna testing; automatic test equipment; calibration; horn antennas; measurement errors; measurement systems; network analysers; polarimetry; OMT polariser; RCS; S parameters measurement; anechoic chamber; automatic vector network analyzer test set; calibration procedure; error model; experimental results; horn antenna; impedance mismatches; leakage; nonsymmetrical radiation patterns; polarimetric measurement systems; wave reflections; Anechoic chambers; Antenna measurements; Antenna radiation patterns; Automatic testing; Calibration; Horn antennas; Impedance; Polarization; Reflection; System testing;
Conference_Titel :
Antennas and Propagation Society International Symposium, 1996. AP-S. Digest
Conference_Location :
Baltimore, MD, USA
Print_ISBN :
0-7803-3216-4
DOI :
10.1109/APS.1996.549577