Abstract
A new high-sensitivity package-leak testing method has been developed for micro electro mechanical system (MEMS) sensors such as a vibrating angular-rate sensor housed in a vacuum package. Procedures of the method to obtain high leak-rate resolution are as follows. (1) A package filled with helium gas is kept in a small and closed chamber to accumulate helium gas leaking out of the package. (2) The chamber is chilled with liquid-N2 after the accumulation to reduce background gases except the helium gas. (3) The helium gas is transported to a mass spectrometer in a short time and recorded as a transient waveform. The leak-rate resolution of 1times10-17 Pamiddotm3/s was obtained. This resolution was 105 times superior to that of the conventional method.
Keywords
electronics packaging; helium; mass spectrometers; microsensors; He; MEMS sensor; angular-rate sensor; high-sensitivity package-leak testing method; leak-rate resolution; mass spectrometer; microelectromechanical system sensor; vacuum package; Gases; Helium; Mass spectroscopy; Mechanical sensors; Mechanical systems; Micromechanical devices; Packaging; Sensor systems; System testing; Vacuum systems;
fLanguage
English
Publisher
ieee
Conference_Titel
Micro Electro Mechanical Systems, 2009. MEMS 2009. IEEE 22nd International Conference on
Conference_Location
Sorrento
ISSN
1084-6999
Print_ISBN
978-1-4244-2977-6
Electronic_ISBN
1084-6999
Type
conf
DOI
10.1109/MEMSYS.2009.4805488
Filename
4805488
Link To Document