DocumentCode
3047265
Title
An effective compression scheme for multilead ECG signals containing ectopie ventricular activation
Author
Linnenbank, A.C. ; Kemmelings, J.G.C. ; Muilwijk, S.L.C. ; Peper, A. ; Grimbergen, C.A.
Author_Institution
Medical Physics &. Informatics Dept., and Cardiology Dept., Academic Medical Center, University of Amsterdam, Meibergdreef 15, 1105 AZ, Amsterdam, the Netherlands
Volume
2
fYear
1992
fDate
Oct. 29 1992-Nov. 1 1992
Firstpage
527
Lastpage
528
Abstract
When recording 64 channel high resolution (0.75μV per bit) eletrocardiograms, datatile compression becomes necessary. It is shown that these files can be stored with 5 bits per sample at 500 Hz sampling rate. The preferred method consists of computing the second order derivative and Huffman encoding of the residue. By using this data compression technique, polymorphic ventricular tachycardia recordings which have a larger information content can be compressed to nearly the same volume as sinus rhythm. It is not necessary to find the optimal Huffman coding for every recording. A fixed encoding scheme can be used without significant increase in file length.
Keywords
Electrocardiography; Histograms; Instruments; Optical films; Optical filters; Presses; Strontium;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 1992 14th Annual International Conference of the IEEE
Conference_Location
Paris, France
Print_ISBN
0-7803-0785-2
Electronic_ISBN
0-7803-0816-6
Type
conf
DOI
10.1109/IEMBS.1992.5761092
Filename
5761092
Link To Document