• DocumentCode
    3053722
  • Title

    Testing device for surgical grounding plates

  • Author

    Nessler, N. ; Huter, H. ; Wang, L.

  • Author_Institution
    Institute dor Applied Physics, University of Innsbruck Technikerstr. 25, A-6020 Innsbruck, Austria
  • Volume
    6
  • fYear
    1992
  • fDate
    Oct. 29 1992-Nov. 1 1992
  • Firstpage
    2388
  • Lastpage
    2389
  • Abstract
    This paper reports about a new device for safety test of disposable grounding plates for electrosurgery. Neutral electrodes (grounding plates, dispersive electrodes) must fulfill the requirements of the american AAMI HF-18 standard. Instead of measurements on human volunteers an electronic devive was built to simulate human skin. Its conductive surface is devided into a number of 1 cm2 pads for a seperately measurement of local current density and local temperature rise. A PC-AT is used as control device and for graphical evaluation of electrical and thermal characteristics of the electrode under test.
  • Keywords
    Current measurement; Electrical resistance measurement; Electrodes; Impedance; Impedance measurement; Temperature measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1992 14th Annual International Conference of the IEEE
  • Conference_Location
    Paris, France
  • Print_ISBN
    0-7803-0785-2
  • Electronic_ISBN
    0-7803-0816-6
  • Type

    conf

  • DOI
    10.1109/IEMBS.1992.5761417
  • Filename
    5761417