• DocumentCode
    3055385
  • Title

    Adaptive control of left ventricular assist devices

  • Author

    Mcinnis, B.C. ; Guo, Z.W. ; Lu, P.C. ; Wang, J.C.

  • Author_Institution
    University of Houston-University Park, Houston, Texas
  • fYear
    1983
  • fDate
    - Dec. 1983
  • Firstpage
    1495
  • Lastpage
    1500
  • Abstract
    This paper presents a computer based adaptive control system for left ventricular assist devices consisting of air driven diaphragm pumps. The system will maintain both the aortic and left atrial pressures of the recipient precisely at desired levels. The system design includes a two-input-two-output adaptive control algorithm which determines the values of the control variables which become inputs to a servo-loop controlled pneumatic pressure source. The performance of the system has been demonstrated by in vitro experiments on a mock circulatory system. The results demonstrate that the assist system will maintain the essential blood pressures precisely at desired levels. Furthermore, when major changes in the output of the mock loop left ventricle occur, the algorithms retunes itself and restores the pressures to desired levels within 10 beats. Hence, we conclude that the system has performance characteristics which are highly desirable for clinical applications.
  • Keywords
    Adaptive control; Electrocardiography; TV;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Decision and Control, 1983. The 22nd IEEE Conference on
  • Conference_Location
    San Antonio, TX, USA
  • Type

    conf

  • DOI
    10.1109/CDC.1983.269786
  • Filename
    4047811