• DocumentCode
    3056701
  • Title

    A method for determining insensible water loss from premature infants by psychrometric techniques

  • Author

    Perry, Edward H. ; Stephens, Carol E. ; Fitch, Charles W. ; Korones, Sheldon B.

  • Author_Institution
    Mechanical Engineering Department, Memphis State University, Memphis, TN 38152
  • Volume
    6
  • fYear
    1992
  • fDate
    Oct. 29 1992-Nov. 1 1992
  • Firstpage
    2508
  • Lastpage
    2510
  • Abstract
    Described is a new method employing psychrometric principles for determining total insensible water loss from premature infants. The method, based on the concept of mass conservation, requires only simple measurements of relative humidity and temperature and thus lends itself well to the role of a clinical technique. Presented is a brief mathematical analysis of the method along with results obtained from in vitro and in vivo comparisons of the new method to the weight-loss method used in the past.
  • Keywords
    Argon; Heating; Plastic films; Weight measurement;